TCP312 賣TCP312探頭,TCP312 賣TCP312探頭,

的詳細資料
The TCP300 and TCP400 Series AC/DC current measurement family is a highly advanced current measurement system for today's current measurement needs. When connected to Tektronix oscilloscopes with TEKPROBE Level II, TekConnect (w/ TCA-BNC), or TekVPI (w/ TPA-BNC) interfaces, current measurements and calculations are simple and easy.

KEY PERFORMANCE SPECIFICATIONS
AC/DC measurement capabilities

- DC - 100 MHz, Current Probe Amplifier (TCPA300) uses:
- DC - 100 MHz, 30 A DC (TCP312A)
- DC - 50 MHz, 50 A DC (TCP305A)
- DC - 15 MHz, 150 A DC (TCP303)
- DC - 50 MHz, Current Probe Amplifier (TCPA400) Uses:< >DC - 2 MHz, 750 A DC1 (TCP404XL) (500 A DC Continuous)眼睛看見的是價格,眼光看見的是價值,同樣的經營模式,不一樣的服務,同樣的買賣儀器,但是我們給你不一樣的享受,以及由此而收獲到的加倍價值,選擇決定命運,眼光決定未來!相信你是那個眼光獨到的人,我們期待你的光臨哦!
:小歐: 2631627011
KEY FEATURES
- Automatic scaling and units2 - Oscilloscope on-screen readout of magnitude and amps reduces measurement errors with no more hand calculations
- AC/DC input coupling
- Low insertion impedance reduces device under test loading
- Split-core construction allows easy circuit connection
- Status indicators provide visual operating status and notification of potential error conditions - degauss, probe open, overload, not terminated into 50 Ω, noncompatible probe type
- Low DC drift and noise allows improved low-level current measurements
- 3rd party safety certification
2 Requires a TDS TEKSCOPE oscilloscope or a TekConect oscilloscope with TCA-BNC adapter

APPLICATIONS
Development and analysis solutions for designers, installers, and service personnel in ecom, data comm, computer, and semiconductor power electronics environments for:
< >Power supplies (switching and linear)Semiconductor devices (SCRs, IGBTs, MOSFETs, CMOS, BJTs)Power inverters/convertersElectronic ballastsIndustrial/consumer electronics communications (phone, salite, relay stations)Motor drivesTransportation systems (electronic vehicles, electric trains, locomotives, avionics)
Meets today's AC/DC current measurement applications
Enter a short description of your concept here (optional).
The TCPA300 amplifier, when used with TCP312A, TCP305A, or TCP303 probes, provides a wide range of current measurement capability and spans the gap between low-level milliamp measurements to very high current levels. These three probes provide current measurement capabilities of 30 A, 50 A, and 150 A DC continuous. For even higher current levels, the TCPA400 amplifier with the TCP404XL current probe measures 500 A DC continuous and 750 A DC continuous, derated with duty cycle.
Higher-frequency performance is available with the TCP312A w/TCPA300 providing ≥100 MHz bandwidth and a maximum current of 30 A DC.
Measurement errors and manual calculations are now a thing of the past
Enter a short description of your concept here (optional).
With this new series of current measurement tools, automatic control and on-screen scaling and units is provided for users of Tektronix TDS3000, TDS500, TDS600, TDS700, TDS5000, TDS6000, and TDS7000B series oscilloscope systems (the DPO3000, MDO/MSO/DPO4000, MSO/DPO5000, and DPO7000 series oscilloscopes, the TPA-BNC adapter is required).
The TCP300/TCP400 current measurement systems seamlessly integrate with your TDS series oscilloscope.
Even non-TEKPROBE systems can use the TCPA300/400 series to make proper current measurements by simply multiplying the measured output voltage on the oscilloscope by the TCPA300/400 series range setting.
| TCP312A w/ TCPA300 | TCP305A w/ TCPA300 | TCP303 w/ TCPA300 | TCP404XL w/ TCPA400 | |
|---|---|---|---|---|
| Bandwidth | DC – 100 MHz | DC – 50 MHz | DC – 15 MHz | DC – 2 MHz |
| Rise time | ≤3.5 ns | ≤7 ns | ≤23 ns | ≤175 ns |
| DC accuracy | ±3% of reading | ±3% of reading | ±3% of reading | ±3% of reading |
| DC accuracy, typical | ±1% of reading | ±1% of reading | ±1% of reading | ±1% of reading |
| Lowest measurable current (at ±3% accuracy at DC) Scope set to 1 mV/div and 20 MHz BW limited | 1 mA | 5 mA | 5 mA | 1 A |
| Maximum Amp-Second product, typical (Based on amplifier range setting) | 50 AμS – 1 A/V 500 AμS – 10 A/V | 500 AμS – 5 A/V NA – 10 A/V | 3,000 AμS – 5 A/V 15,000 AμS – 50 A/V | NA – 1 A/mV |
| Maximum wire voltage, bare insulated | 150 V CAT II 300 V CAT II | 150 V CAT II 300 V CAT II | 600 V CAT I & II 300 V CAT III | 600 V CAT I & II 300 V CAT III |
| AC-coupling low-frequency Bandwidth, typical (Low pass – 3 dB point) | <7 Hz | <7 Hz | <7 Hz | <7 Hz |
| Displayed RMS noise, typical (at 20 MHz bandwidth limit) | ≤250 μARMS | ≤1.25 mA RMS | ≤2.5 mARMS | ≤250 mARMS |
| Signal delay (to output BNC) | 17 ns | 19 ns | 40 ns | 80 ns |
| Insertion impedance | 0.11 Ω at 1 MHz 0.12 Ω at 10 MHz 0.35 Ω at 50 MHz 0.7 Ω at 100 MHz | 0.02 Ω at 1 MHz 0.1 Ω at 10 MHz 0.35 Ω at 50 MHz | 0.01 Ω at 1 MHz 0.025 Ω at 5 MHz 0.1 Ω at 15 MHz | 0.1 m Ω at 10 kHz 0.6 m Ω at 100 kHz 8 m Ω at 1 MHz 16 m Ω at 2 MHz |
CHARACTERISTICS
Maximum current ratingsHigh-current sensitivity
| TCP312A w/ TCPA300 | TCP305A w/ TCPA300 | TCP303 w/ TCPA300 | TCP404XL w/ TCPA400 | |
|---|---|---|---|---|
| Range | 10 A/V | 10 A/V | 50 A/V | 1 A/mV |
| DC (continuous) | 30 A | 50 A | 150 A | 500 A (750 A) |
| RMS (sinusoidal) | 21.2 A | 35.4 A | 150 A | 500 A |
| Peak | 50 A | 50 A | 500 A | 750 A |
Low-current sensitivity
| TCP312A w/ TCPA300 | TCP305A w/ TCPA300 | TCP303 w/ TCPA300 | TCP404XL w/ TCPA400 | |
|---|---|---|---|---|
| Range | 1 A/V | 5 A/V | 5 A/V | N/A |
| DC (continuous) | 5 A | 25 A | 25 A | N/A |
| RMS (sinusoidal) | 3.5 A | 17.7 A | 17.7 A | N/A |
| Peak | 50 A | 50 A | 500 A | N/A |
PHYSICAL CHARACTERISTICS